6-8
6. Troubleshooting
<Patient data inquiry to the TOMEY Link>
z
z
z
z
z
"Comm. Busy!! Retry!!" is displayed.
Cause 1
Because of the internal communications, communi-
cations are not received temporarily.
Remedy
Give operation again after a short time of
elapse.
z
z
z
z
z
"NO ID!!" is displayed.
Cause 1
Patient ID number has not been inputted.
Remedy
Give communication operation after inputting
the patient ID number.
z
z
z
z
z
"LAN Connection ERROR!!" is displayed.
Cause 1
The connection of the LAN cable is not correct.
Remedy
Check to see if the connection of the LAN
cable to the connector provided at the rear
side of the instrument unit and to the net-
work side (the switching hub) is proper. Es-
pecially as for the connection of the LAN
cable to the network side, ask your system
administrator for correct connection.
Cause 2
The power of the hub is not on.
Remedy
Check if the power of the hub wher LAN cable
is connected is on.
z
z
z
z
z
"Server Timeout!!" is displayed.
Cause 1
No server's response is caused by busy condition
of the TOMEY Link server.
Remedy
Give operation again after a short time of
elapse.
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...