4-11
4.3.5
HAIGIS optimized / HAIGIS Standard
1.
Implanted IOL power (D)
Where,
d=a0+a1 ·ACD+a2 ·L
(ACD
g
0)
d=(a0-0.241·a1)+(a2+0.139·a1)L
(ACD = 0)
a0=0.62467A-72.434 -------*
2.
Post surgery expected refractive power (D)
Where,
X=d2·P+1000L·na-d·L·P
Y=na(1000·na - L·P+d·P)
na:
Refractive index of aqueous and vitreous body
=1.336
nc:
Corneal refractive index =1.3315
A:
A-Constant
RC:
Average corneal curvature (mm)
=(K1+K2)/2
DC:
Average corneal refractive power (D)
L:
Axial length (mm)
ACD: Anterior chamber depth (mm)
REF: Post surgery expected refractive power (D)
V:
Vertex distance (mm) = 12
P:
Implanted IOL power (D)
a1:
0.4 -------*
a2:
0.1 -------*
* The HAIGIS optimized formula calculates with the use of
registered a0, a1,and a2.
1000na na
P
= ———— - ———————
L - D
na
d
—— - ———
Z 1000
REF
Z
= DC + ———————
REF·V
1 - ———
1000
1000(nc - 1)
DC= —————
RC
1000(1000Y-DC·X)
REF
iol
= ————————————
V(1000Y - DC ·X) + 1000X
4.3 IOL power calculation