4-6
Ofst:
Calculated distance between iris and optical surface
of the implanted IOL (including corneal thickness)
(mm)
= ACDconst-3.336
= (0.62467A-68.747)-3.336
H:
Height of the corneal dome (mm)
= r - r
2
-W
2
/4
A:
A-Constant
K:
Average corneal refractive power (D)
= (K1+K2)/2
P:
Implanted IOL power (D)
V:
Vertex distance (mm) = 12
na:
Refractive index of aqueous humor and vitreous body
= 1.336
nc:
Corneal refractive index = 1.333