8-1
8. SPECIFICATIONS
8.1 Specifications
8.1.1
B-scan Diagnosis
z
Focus
Dynamic focus
z
Frame rate
High:
22 frames/sec.
Low:
11 frames/sec.
z
Maximum moving images (Assistant Function of Biometry)
202 images
z
Maximum waveform data taking number (A-scan diagnosis function)
202 images
z
Image display range
Normal:
34.5mm
×
46.4°
(Ultrasound velocity: 1,550m/s)
Wide:
46mm
×
46.4°
(Ultrasound velocity: 1,550m/s)
z
Vector-A mode
Display step:
0.4°
z
Color scale
240 gradation sequences
z
Distance measurement
<Distance accuracy between cursors>
±
0.5mm
<Minimum distance unit between cursors>
Normal:
0.075mm step (V = 1,550m/s)
(Displayed by rounding off)
Wide:
0.10mm step (V = 1,550m/s)
z
Area calculation
Accuracy:
±
S (mm
2
), where S is calculated area
8. Specifications