6-19
6. Troubleshooting
z
z
z
z
z
The saved image loading function cannot be used.
Cause 1
The data was not saved in the Echo Data mode.
The memory card is not saved with data.
Remedy
The data, loaded on main unit, is only avail-
able in the memory card with the Echo Data
format. The data saved by JPEG format can-
not be loaded. Before using this function,
please make sure that the images are stored
in memory card by Echo Data format.
z
z
z
z
z
TOTAL/CONTRAST volume cannot work.
Cause 1
The volume has not selected as TOTAL/CON-
TRAST.
Remedy
Press the Volume Function Select button-1
to select the function to TOTAL/CONTRAST.
z
z
z
z
z
NEAR/FAR volume cannot work.
Cause 1
The volume function has not selected as NEAR/
FAR.
Remedy
Press the Volume Function Select button-2
to select the function to NEAR/FAR.
Cause 2
The unit is not in scanning, but in “Freeze” mode.
Remedy
The unit should be in scanning mode in or-
der to change NEAR/FAR setting.
z
z
z
z
z
"Probe Error" comes up on screen, when switching to the assistant
function of biometry.
Cause 1
The connection of the B-scan probe was not cor-
rect.
Remedy
Make sure that the B-scan probe is inserted
sufficiently.
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...