3-107
<Chin measurement>
Continue measurement until all ten pieces of measured data
come within the range of ±0.1mm to the average. It is also
noted that this measurement can be discontinued on the way of
measurement.
1) The instrument reports it with a monitor sound of "beep"
when the measuring conditions are satisfied.
2) The instrument automatically starts measuring data when
measurement data becomes stable.
When the instrument completes taking data, it reports with
a sound of "beep".
3) When 10 pieces of measured data have been taken into the
instrument, it finishes measurement with a reporting sound
of "beep".
The waveforms being displayed in the monitor screen are
those for measured data which are closest to the average of
the data.
b) Manual measurement (Manual)
Ԙ
24+06
#$
616#.
%1064#56
(#4
$KQ
(4''<'
0'#4
(4''<'
5#8'24+06
1) The instrument reports it with a monitor sound of "beep"
when the measuring conditions are satisfied.
2) Press the FREEZE button
A
or the FREEZE pedal for
footswitch
B
to take measured date in the instrument.
3) Again press the FREEZE button
A
or the FREEZE pedal
for footswitch
B
to proceed to the next measurement.
4) Take measurement in a similar manner after above until 10
pieces of data are taken with a sound of "beep" for report-
ing the finishing of measurement.
(Fig. 1)
(Fig. 2)
3.6 Biometry
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...