6-4
6. Troubleshooting
Cause 3
The setting of print mode in the axial length mea-
suring function is wrong.
Remedy
Set the printing mode at "Standard" or
"Simple" in the utility of the axial length mea-
suring function.
z
z
z
z
z
The following error messages were displayed.
“DO NOT EXIST,TP_CAL_DATA”
“INCORRECT,TP_CAL_DATA”
“WRITE ERROR,TP_CAL_DATA”
Cause 1
Nonconforming touch panel
Remedy
After the above messages are displayed for
5 seconds, the instrument changes the
screen mode to the Calibration screen. Touch
the center of the white "+" marking in the
black screen. When touching said center, the
point disappears and shows the next "+"
marking. Touch the center of the marking in
the same manner until the "+" marking will
not be displayed. When this operation is fin-
ished, the initial screen of the B-scan image
diagnosis function is displayed to make the
ordinary operation possible. After the above
operation is made, the instrument will con-
tinue operation without problem until the
power source s turned off. If the same error
messages are followed every time when turn-
ing the power on, ask your Tomey represen-
tative for inspection and repair.
z
z
z
z
z
The button which is different from that pressed in the touch panel
was reacted.
Cause 1
The calibration for touch panel was shifted.
Remedy
Turn the power source on while pressing the
PRINT button and the A/B-Bio select button.
If this state elapses for 5 seconds, the screen
will be changed to the calibration screen for
touch panel. Then give calibration as
specifed.
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...