6-22
6. Troubleshooting
Cause 2
Ambient noises
Remedy
If there is any noise generation source (such
as motor or laser surgery equipment) located
close to this instrument, move such noise
source away from the instrument.
Cause 3
The fixation light for patient is not proper.
Remedy
Guide the patient's eye direction by using the
fixation light on chin rest.
Cause 4
The eye contacting part of the biometry probe is
damaged.
Remedy
When the probe is damaged, promptly stop
using it and consult your local representa-
tive for countermeasures.
z
z
z
z
z
Hand mode (the hand holding probe) cannot work well.
Cause 1
In case of the hand holding probe, it’s hard to meet
all measurement conditions without any clue or ex-
perience.
Remedy
It is not necessary to continue to take data in
until automatic freeze is reached. If measure-
ment is difficult, you can stop measurement
and print or save data.
Remedy
For stable measurement, have the patient lay
himself, which helps you hold the probe
stable and control the probe angle. It is also
recommended to take measurement in such
a manner as applying the probe weight
against the probe, which does not harm the
eye.
Remedy
Lightly touch your hand holding the probe to
the patient forehead or cheek to secure the
position and posture of your probe holding
hand.
z
z
z
z
z
Manual measurement cannot work.
Cause 1
The connection of the footswitch is not correct.
Remedy
Connect the footswitch properly so that it is
securely locked, in accordance with "3.2.1 d)
Connection of the footswitch".
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...