3-68
1) Press the cursor line moving key (UP/DOWN)
A
, so the cur-
sor line will move up and down. Move the cursor line, so each
of the cornea front cursor
B
, the crystalline lens front cursor
C
, the crystalline lens rear cursor
D
, and the retina cursor will
automatically move to its automatically detected position. If,
however, not automatically detects, the cursor will not be dis-
played. The moving range of the cursor line is ±10° from the
center axis of the probe.
2) Press the SW-C key
F
, so the movable cursor (red: active
cursor) will be changed to the other mode.
3) Press the cursor moving key (left/right)
G
, so the active
cursor will move left and right over the cursor line.
4) The screen then shows the distance
H
between the cornea
front cursor and the retinal cursor and that
I
between the
cornea front cursor and the crystalline lens front cursor. As
for setting of the sound velocity, see "3.4.8 a) Eye type
setting function".
e) Echo data saving
As for the operating procedure, see "3.3.8 i) Echo data save
function" of the B-scan image diagnosis function. The saving
data include set values of images, ID number, patient's name,
the eye to be examined, patient's gender, physician's name, probe
applying angle, data and time, measuring results, each cursor
and cursor line position, contrast (CT), total gain (TG), near
gain (NG), and far gain (FG).
f) JPEG image saving
See "3.3.8 j) JPEG image saving".
g) Transfer of patient data
See "3.3.8 l) Transfer of patient data".
3.4 Assistant function of Biometry
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...