6-13
6. Troubleshooting
z
z
z
z
z
"Registration Error!! (***) " appears.
Cause 1
TOMEY Link server/ Conputer of DATA Transfer
has temporary memory shortage .
Remedy
Send it again. If this error appears often, in-
crease RAM of the computer.
Cause 2
Software version is not correct.
Remedy
Check the version of the software of the in-
strument, TOMEY Link/DATA Transfer and
contact the place of the purchase.
Cause 3
TOMEYLink server / DATA Transfer is not up-
graded enough to respond the request.
Remedy
TOMEYLink Server software should be up-
graded into the latest version. Contact your
local TOMEY Representatives.
z
z
z
z
z
The following error messages come up.
"Data ERROR!!(***)"
"SYSTEM ERROR!!(***)"
Cause 1
These error messages are caused by internal
nonconformities.
Remedy
Try again minutes later. If this error often
appears, contact the place of purchase.