6-11
6. Troubleshooting
<Patient data transfer>
z
z
z
z
z
The SEND button is not available.
Cause 1
"CONNECT" is not selected in communication
setting.
Remedy
Make correct setting in "3.8.2 Setting items
for B-scan diagnosis assistant function/A-
scan diagnosis function f) Setting of data
communication"
z
z
z
z
z
"Comm. Busy!!" is displayed.
Cause 1
The communication cannot be established tempo-
rarily because of internal processes.
Remedy
Try again after a while.
z
z
z
z
z
"NO ID!!" is displayed.
Cause 1
Patient ID number is not input.
Remedy
Try again after inputting the patient ID num-
ber.
Cause 2
Sending without ID is not set allowed.
Remedy
Make correct setting in "3.11.2 Setting items
on Utility display f) Setting of data communi-
cation".
z
z
z
z
z
"OD/OS is not selected!!" is displayed.
Cause 1
The eye to be measured is not selected.
Remedy
The eye to be measured is not selected.
Choose either OD or OS for measured data
in order to send it to TOMEYLink.