3-93
3.6.3
Setting of measuring conditions
The following conditions are set immediately after turning the power on
and after switching the function to the biometry mode.
· Eye to be measured:
Right eye (R)
· Type of eye to be measured:
Normal (Sound velocity to be same
as set in the previous use.)
· Measuring method:
Same as that in the previous use
· Gain:
Same as that in the previous use
· Measuring mode:
Same as that set in the previous use
· Axial length calculation method:
Same as that set for the previous use.
For measuring the new patient, be sure to delete all previ-
ous patient data before pressing the NEW key before car-
rying out measurement for the new patient (See "3.6.9 a)
Deleting of measurement data for both eyes (Measuring
preparation for the new patient)". If taking measurement
without pressing the NEW key, measured data may other-
wise be mixed with data for previous patient.
a) Setting of the eye to be measured
)CKP
#XGTCIG#:+#.
5&
OO
OO
8GNQEKV[#XI
8GNQEKV[.'05
8GNQEKV[#%&
OO
*CPF
0QTOCN
M/V
0Q
#XI
#:+#.
#%&
.'05
+&0Q
5GZ/CNG
#WI
2CVKGPV61/';
2J[UKEKCP;#/#&#
OU
OU
OU
#:+#.
+1.
7VKNKV[
'FKV
4CPIG
R
1
&
4GVCMG
0'9
4GVKPC
%WTUQT
The right eye and the left eye are alternately displayed every
time when pressing the Right/Left key
A
. The biometry func-
tion and the IOL power calculation function (See "3.7 IOL power
calculation".) acquire and save measured data for the right and
left eye respectively. Accordingly, be sure to set the conditions
for the eye to be measured before calculation.
A
3.6 Biometry
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...