3-10
3.2.3
How to connect the instrument to the Chin Rest AL-1100
For biometry measurement using the Chin Rest AL-1100, con-
nect the supplied Chin Rest AL-1100 (option device) in the fol-
lowing procedure.
a) How to connect the biometry probe
Make sure that the slider part moves smoothly.
1) Carefully insert the biometry probe
A
from the side of the
slider operator. (Fig. 1) Care must be taken not to damage
the eye contacting part of the probe.
2) Bundle the probe cord double and hold it with the cord
hook
B
of the Chin Rest. Twenty centimeters from the probe
to the cord hook allows measurement easy, without caus-
ing the slider is forcedly pulled with the measuring probe.
(Fig. 2)
b) How to connect the power plug of Fixation Lamp for Chin Rest, AL-1100
Insert the fixation lamp power plug
A
of the Chin Rest into the
terminal
B
having the identification of "FIX LIGHT".
Ԙ
OO
ԙ
.#0
8+&'1176
24+06'4
25
%
/
(+:.+)*6
4'/16'
/
.+0-
(Fig. 1)
(Fig. 2)