8-10
Appendix
Electromagnetic guidance
8. Appendix
Caution: Medical electrical equipment.
EMC (Electro Magnetic Compatibility) must be considered before any medical electrical equipment is
installed or put into service. Follow the information in the accompanying documentation when installing
and operating the
UD-6000.
Caution: Portable or mobile RF communication equipment can effect Medical Electrical equipment.
Guidance and manufacturer's declaration
electromagnetic emissions
Table 201
The
UD-6000
is intended for use in the electromagnetic environment specified below.
The customer or the user of the
UD-6000
should assure that it is used in such an environment.
Electromagnetic environment - guidance
Emissions test
Compliance
RF emissions
CISPR 11
Group 1
The
UD-6000
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The
UD-6000
is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...