6-7
6. Troubleshooting
<ID number input from external device>
z
z
z
z
z
Even when a Patient ID is input by external device, the beep sounds
without showing anything on screen.
Cause 1
The PS/2 port connection is not correct.
Remedy
Correctly connect the PS/2 port of the exter-
nal connection ID input device to its corre-
sponding connector located at the rear side
of the main body of the instrument.
Cause 2
The screen which does not make ID input is being
displayed.
Remedy
The ID inputting from the external connec-
tion ID input device can only be made when
displaying the screen mentioned in "3.3.5 In-
putting of the ID number from the external
connection ID input device (barcode reader,
etc.)". Make ID input after these screens have
been displayed.
z
z
z
z
z
"Illegal Code !!" appears.
Cause 1
The letters to which the instrument do not respond
were inputted from the external connection ID
input device.
Remedy
The letters which can be handled with this
instrument are half alphameric characters (0
to9, A to Z, a to z), part of symbols (., *, /, +, -
, BS, ENTER) only. If any letters other than
these letters, no display can be given. If any
letters other than these letters are mixed on
the way of letter strings, only the letters that
can be displayed will be displayd.
Summary of Contents for UD-600
Page 2: ......
Page 14: ...xii This page is intentionally blank...
Page 22: ...1 8 This page is intentionally blank...
Page 28: ...2 6 This page is intentionally blank...
Page 157: ...3 129 Simple mode 3 9 Printout...
Page 159: ...3 131 Simple mode 3 9 Printout...
Page 192: ...3 164 This page is intentionally blank...
Page 256: ...8 14 This page is intentionally blank...
Page 260: ...A 4 INDEX...
Page 262: ...0123 ID 0608...