CombiFlash® Rf
+
User Manual
1-16
Peak Detection Modes
Slope, Threshold, or Time Windows
Flow Cell Path Length
.1 mm, ±.025mm. Optional .2mm, .3mm, .5mm, 1mm,
and 2mm.
UV Detection and
Monitor Wavelength
Selectable from 200 to 400 nm range. Optional 200 to
800 nm UV-Vis detector.
Wavelength Accuracy
±5 nm
Fraction Accuracy
±6%
ELSD Options Only
Gas Inlet Pressure
60 to 70 psig
Gas Consumption
<2.5 SLPM
Drift Tube Temperature
Setting range: 30 to 90 °C
Split Flow Rate
0.75 mL/min, ±10%
Spray Chamber
Setting range: 10 to 60 °C
Temperature limited to minimum of 5 °C below ambient
Electrical Safety per EN 61010-1
Pollution Degree
2
Installation Category
II
Maximum Altitude
2000 meters
a. All specifications are subject to change.
b. For flash columns undetected by RFID, the max pressure is limited to 50 psi.
For columns with a RIFD tag, pressure limit is determined by column pres-
sure rating.
Table 1-5 CombiFlash EZ Prep Specifications
a
(Continued)