2.4 Connect Solvent Lines (Excludes EZ Prep)
WARNING
Risk of fire ignited by electrostatic discharge. Never substitute
the black tubing on Combi
Flash
systems. The black tubing
(P/N 023-0503-06) is conductive. This tubing is required to
dissipate static electricity.
Figure 2-2 Recommended solvent connection
60-0923-017 Ferrule
60-0923-015
Compression Fitting Nut
60-5235-050 38 mm Cap, or
60-5235-051 45 mm Cap
60-5234-129
Inlet Filter Assembly
(Place weighted filter at
bottom of container.)
To Solvent Level Sensing
To Solvent Supply
Connection