CombiFlash® Rf
+
User Manual
8-22
A plug has occurred in
the PurIon fluid lines
.
The plug must be
corrected to allow
continued operation.
Error 316.
This error occurs if the tubing is plugged while the
method development screen is in use. See cleaning
the probe capillary.
The ionization probe
(ESI or APCI) isn’t
fully seated into the
ion source housing.
Please ensure the
probe is fully seated,
then press O
K
to
continue. Error 317.
On PurIon systems (not PurIon S or PurIon L), the
probe isn’t seated properly. Loosen the
thumbscrew on the front of the ion source housing,
push down on the probe, then tighten the
thumbscrew.
The fluid interface
pressure is too low.
Error 325
This could be due to lack of carrier fluid, loss of
pump prime, or leakage
.
Table 8-3 General Troubleshooting
Error/Symptom
Solution
No signal on any
mass, or very weak
signal
• Verify nitrogen pressure. Nitrogen is required
to nebulize the sample.
• Check carrier solvent level, refill carrier.
solvent and re-prime the carrier solvent
pump. Carrier solvent is needed to bring
sample to the mass spectrometer.
• Check that the priming valve is tight. A loose
priming valve prevents sample from getting
to the mass spectrometer.
• Verify the splitting valve is not leaking. This
causes loss of carrier solvent preventing
sample from getting to the mass
spectrometer.
• Check for leaks around other fittings on the
solvent interface
Table 8-2 Common PurIon Error Codes and Resolutions
Summary of Contents for CombiFlash EZ Prep
Page 2: ...CombiFlash Rf User Manual ii ...
Page 8: ...CombiFlash Rf User Manual viii ...
Page 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Page 85: ...CombiFlash Rf User Manual 3 20 ...
Page 103: ...CombiFlash Rf User Manual 4 18 ...
Page 185: ...CombiFlash Rf User Manual 6 64 ...
Page 199: ...CombiFlash Rf User Manual 7 14 ...
Page 223: ...CombiFlash Rf User Manual 8 24 ...
Page 225: ......
Page 227: ......
Page 229: ......
Page 231: ......
Page 233: ......
Page 235: ......
Page 237: ......