CombiFlash® Rf
+
User Manual
6-32
particular compound by reducing fragmentation or adduct
formation.
•
Mass Spectrum Graph
- this displays the mass
spectrum. There are two mass spectra displayed. Blue
displays the maximum of the current injection. Red
displays the maximum of the previous injection for
comparison. When a new injection is performed, the blue
spectrum replaces the red spectrum and a new blue
spectrum is started. To inject a sample, move the PurIon
injection valve to the “Inject Sample” position.
Inject the sample (20 uL) and move the valve to the
“Scan Mass” position.
Note
Note:
Only
use
22
gauge
square
tip
needles
(PN
29
‐
9001
‐
911)
to
avoid
scratching
the
internal
surfaces
of
the
injection
valve.
Filter
the
sample
with
a
0.45
µ
filter
to
avoid
clogging
the
PurIon
tubing
and
probe
capillary.
Sample
concentration
should
be
less
than
20
µg/mL.
•
POLARITY
- switches the mass spectrometer between
positive and negative polarity.
•
FORMAT GRAPH
- Set the displayed range of the
mass spectrum. The PurIon system still collects the
entire mass range even when the range is set within
than the minimum and maximum values allowed for the
spectrometer. Values can range from 50 to 1200 Da
(PURION and PURION S) or 2000 Da (PURION L).
Changing the mass range defined by these controls may
cause the Y-axis (ion count) scaling to change based on
the tallest peak within the range.
•
Fluid Interface Pressure
- The carrier solvent
pressure is displayed in a ribbon gauge. If the pressure
approaches the level of the red line, the system is
clogged and should be cleaned to allow proper operation.
•
ION SETTINGS
- Allows selection of different ion
settings to maximize the intensity of the desired
molecular ion. The factory selections include “Robust”
for compounds that do not easily ionize. “Typical” works
well for most compounds. The “Fragile” setting is used
Summary of Contents for CombiFlash EZ Prep
Page 2: ...CombiFlash Rf User Manual ii ...
Page 8: ...CombiFlash Rf User Manual viii ...
Page 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Page 85: ...CombiFlash Rf User Manual 3 20 ...
Page 103: ...CombiFlash Rf User Manual 4 18 ...
Page 185: ...CombiFlash Rf User Manual 6 64 ...
Page 199: ...CombiFlash Rf User Manual 7 14 ...
Page 223: ...CombiFlash Rf User Manual 8 24 ...
Page 225: ......
Page 227: ......
Page 229: ......
Page 231: ......
Page 233: ......
Page 235: ......
Page 237: ......