Section 8 Maintenance
8-19
with three small holes, are facing upwards towards the
outlet. See Figure 8-12.
Figure 8-12 Correct orientation of cartridge
2. Tighten check valve holders finger tight, then with pliers
tighten an additional ¼ turn.
8.7 Injection Valve Maintenance
Combi
Flash
Rf
+
systems, including those equipped with 4x,
have an automatic injection valve. Under normal operating
conditions, this self-cleaning valve does not require any
maintenance.
If maintenance is required, refer to Instruction Sheet
60-5232-475, Combi
Flash
Rf Injection Valve Replacement.
8.8 Electrical Troubleshooting
If your instrument stops working and the touch panel display
is off, check the line cord connection.
If the line cord is connected properly, check the fuses in the
power input module located on the system’s rear panel. These
are time delay fuses (T-type) and rated according to the mains
input power voltage.
Combiflash Rf
+
, Rf
+
Purion, and Rf 150 systems:
• 100 to 129 VAC mains input: 2.0 A fuse, one required
• 215 to 264 VAC mains input: 1.0 A fuse, two required
Combiflash Rf
+
Lumen system:
• 100 to 129 VAC mains input: 3.0 A fuse, one required
• 215 to 264 VAC mains input: 3.0 A fuse, two required
Summary of Contents for CombiFlash EZ Prep
Page 2: ...CombiFlash Rf User Manual ii ...
Page 8: ...CombiFlash Rf User Manual viii ...
Page 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Page 85: ...CombiFlash Rf User Manual 3 20 ...
Page 103: ...CombiFlash Rf User Manual 4 18 ...
Page 185: ...CombiFlash Rf User Manual 6 64 ...
Page 199: ...CombiFlash Rf User Manual 7 14 ...
Page 223: ...CombiFlash Rf User Manual 8 24 ...
Page 225: ......
Page 227: ......
Page 229: ......
Page 231: ......
Page 233: ......
Page 235: ......
Page 237: ......