Section 4 Preparation: Optional 4x Module
4-3
4.4 Connect Solvent Lines
WARNING
Risk of fire ignited by electrostatic discharge. Never substitute
the black tubing on Combi
Flash
systems. The black tubing
(P/N 023-0503-06) is conductive. This tubing is required to
dissipate static electricity.
CAUTION
To prevent damage or premature wear to the pump and internal
valves, clean solvent should be used. The solvent should not
contain any dissolved solids.
The Combi
Flash
Rf system has four solvent ports on the left
side panel. To complete the solvent connections:
Figure 4-1 Labeled ports on CombiFlash Rf system side panel
4.4.1 Connecting the Combi
Flash
Rf with 4x Module
1. Remove the jumper tubing from ports A, B, C, and D.
2. Locate the solvent lines on the side of the Rf 4x module.
These lines are labeled A through D. Connect the solvent
lines to their respective ports on the Combi
Flash
Rf
system.
Front