1.4 Specifications
Table 1-1 Combi
Flash
Rf 150 and Rf
+
System
Specifications
1
Rf 150 Systems
(Without Inject Valve)
Rf
+
Systems
(With Inject Valve)
Overall Dimensions
(H
W
D)
61 x 36 x 43 cm (24 x 14.1 x 17")
Weight
22.3 kg (49.2 lbs)
24.5 kg (54 lbs)
Power Options
Input voltage range from 100 to 240 VAC, 50/60 Hz,
2.0 Amps maximum.
Line cord is the disconnect device.
Line Frequency
50/60 Hz
Ambient Temperature
20 to 40 °C (maximum temperature must be at least
15 °C below the boiling point of the solvent)
Humidity (when
connected to power)
95% relative humidity maximum at 20 to 40 °C
Flow Rate Range
10 to 100 mL/min
1 to 200 mL/min
Flow Rate Accuracy
(tested with water at
138 kPa or 20 psi)
±10%
±5% (5-200 mL/min)
Pressure Limit
2
1,034 kPa (150 psi)
Up to 1379 kPa (200 psi)
(column dependent)
Pressure Accuracy
10% of full scale
5% of full scale
Gradient Formation
Binary gradient from 1
and 2 solvent inlets.
Optional manual valves
to select from 1, 2, and 3,
4 inlets.
Binary gradient from 1, 2,
and 3, 4 solvent inlets.
Solvent selection can be
automatically controlled
by the method.
Gradient Accuracy:
±2% of full scale, typical;
±5% maximum
±1% of full scale
Peak Detection Modes
Slope or threshold
Flow Cell Pathlength
0.1 mm, ±25%
Summary of Contents for CombiFlash EZ Prep
Page 2: ...CombiFlash Rf User Manual ii ...
Page 8: ...CombiFlash Rf User Manual viii ...
Page 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Page 85: ...CombiFlash Rf User Manual 3 20 ...
Page 103: ...CombiFlash Rf User Manual 4 18 ...
Page 185: ...CombiFlash Rf User Manual 6 64 ...
Page 199: ...CombiFlash Rf User Manual 7 14 ...
Page 223: ...CombiFlash Rf User Manual 8 24 ...
Page 225: ......
Page 227: ......
Page 229: ......
Page 231: ......
Page 233: ......
Page 235: ......
Page 237: ......