CombiFlash® Rf
+
User Manual
4-2
4.2 Instrument Location
The Combi
Flash
Rf system requires approximately 1550
square centimeters (240 in
2
) of level bench space. Ensure that
the Combi
Flash
Rf has at least 3 cm (1.25") of air space behind
it for ventilation. Additional space may be required for solvent
and waste containers.
The optional 4x expansion module is installed to the left of the
Combi
Flash
Rf system. If paired with the Rf
+
PurIon mass
spectrometer, the 4x expansion module must be installed to
the right of the PurIon. The 4x module requires an additional
933 cm
2
(145 in
2
) of level bench space. Ensure that the module
has at least 3 cm (1.25") of air space behind it for ventilation.
Refer to Table 1-4 for environmental conditions and power
requirements.
Before making any connections to the Combi
Flash
Rf 4x
module, place it on the bench or in the fume hood where it will
be operated. Temporarily position the system so you can
access the back panels to complete the connections.
4.3 Connect Power
Ensure that the On/Standby switch below the Combi
Flash
Rf
system touch screen panel is in the Standby position. Then,
use the supplied IEC power cord to connect the Combi
Flash
Rf
4x module to mains power.
WARNING
Mains power must meet the voltage, frequency, and amperage
requirements listed on the serial number label.
WARNING
As long as the AC mains power cord is connected, power is
inside the unit. The mains power cord is the disconnect device.
Position the Combi
Flash
Rf system and Rf 4x module so that
the power cord can be unplugged, or use a power strip where
the plug can quickly be removed from the outlet in the event
of an emergency.
Summary of Contents for CombiFlash EZ Prep
Page 2: ...CombiFlash Rf User Manual ii ...
Page 8: ...CombiFlash Rf User Manual viii ...
Page 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Page 85: ...CombiFlash Rf User Manual 3 20 ...
Page 103: ...CombiFlash Rf User Manual 4 18 ...
Page 185: ...CombiFlash Rf User Manual 6 64 ...
Page 199: ...CombiFlash Rf User Manual 7 14 ...
Page 223: ...CombiFlash Rf User Manual 8 24 ...
Page 225: ......
Page 227: ......
Page 229: ......
Page 231: ......
Page 233: ......
Page 235: ......
Page 237: ......