Section 6 User Help Reference
6-47
The %B gradient mix and run length are shown on the plot
area. Refer to the D
topic for more
details. The other settings are:
•
Flow Rate
— When using the Isco default column
methods, the flow rate is automatically set to the
optimum flow rate of the column. If using a non-Isco
column, refer to the manufacturer’s column
documentation to determine the best flow rate for the
column. Note that if the column is not detected by the
system the operating pressure is limited to 50 psi.
•
Solvent A and B
— Select the desired solvents from the
lists. The solvents are those defined by the
C
window.
6.4.2.2 Advanced method settings
window gives you access to
advanced settings such as equilibration volume, initial waste,
air purge, peak detection, and peak collection mode.
Equilibration Volume
Column equilibration is optional, but
you may find that separations are more effective if the column
is equilibrated before injecting the sample. The recommended
equilibration volume is at least three column volumes of the
starting mobile phase.
RediSep Silica gel column volumes are:
• RediSep Rf 4 g = 4.8 mL
• RediSep Rf 12 g = 16.8 mL
• RediSep Rf 24 g = 33 mL
• RediSep Rf 40 g = 48 mL
• RediSep Rf 80 g = 125 mL
• RediSep Rf 120 g = 192 mL
• RediSep Rf 220 g = 334 mL
• RediSep Rf 330 g = 443 mL
Other media types will have similar column volumes. Refer to
the R
for actual
volumes.
Summary of Contents for CombiFlash EZ Prep
Page 2: ...CombiFlash Rf User Manual ii ...
Page 8: ...CombiFlash Rf User Manual viii ...
Page 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Page 85: ...CombiFlash Rf User Manual 3 20 ...
Page 103: ...CombiFlash Rf User Manual 4 18 ...
Page 185: ...CombiFlash Rf User Manual 6 64 ...
Page 199: ...CombiFlash Rf User Manual 7 14 ...
Page 223: ...CombiFlash Rf User Manual 8 24 ...
Page 225: ......
Page 227: ......
Page 229: ......
Page 231: ......
Page 233: ......
Page 235: ......
Page 237: ......