ACCQPrep HP125
Section 5 PeakTrak
5-22
To prime the pump, open the priming valve on the fluid
interface counterclockwise. Use the 5 mL Luer lock
syringe provided with the accessory kit to draw liquid
into the pump. Repeat this procedure once more to
completely fill or purge the line from the carrier solvent
reservoir to the priming valve. Make sure the priming
valve is tightly shut after priming the system.
• Stop Carrier Pump- Stops the carrier pump. This
changes to read START CARRIER PUMP after the
pump is stopped.
• Operate Splitter Valve- Runs the splitter valve to verify
operation.
• Fluid Interface Pressure - The carrier solvent pressure
is displayed in a ribbon gauge. If the pressure
approaches the level of the red line, the system is
clogged and should be cleaned to allow good response.
• Close - Closes the window and stops the carrier pump if
it is running.
5.2.13 Ionization Settings
The MS->I
ONIZATION
S
ETTINGS
menu command allows a user to
adjust various settings to improve the ionization of a particular
compound by reducing fragmentation or adduct formation.
• Mass Spectrum Graph - this displays the mass
spectrum. There are two mass spectra displayed. Blue
displays the maximum of the current injection. Red
displays the maximum of the previous injection for
comparison. When a new injection is performed, the blue
spectrum replaces the red spectrum and a new blue
spectrum is started. To inject a sample, move the PurIon
injection valve to the “Inject Sample” position.
Inject the sample (20 uL) and move the valve to the
“Scan Mass” position.
Note
Only use 22 gauge square tip needles (PN 29-9001-911) to
avoid scratching the internal surfaces of the injection valve. Fil-
ter the sample with a 0.45 µ filter to avoid clogging the PurIon
tubing and probe capillary. Sample concentration should be
less than 20 µg/mL.
• Polarity - switches the mass spectrometer between
positive and negative polarity.
Note
T
he PurIon S and L systems capture both polarities. The setting
changes which polarity is displayed.
• Format Graph- Set the displayed range of the mass
spectrum. The PurIon system still collects the entire
Summary of Contents for ACCQPrep HP125
Page 2: ......
Page 6: ...ACCQPrep HP125 Installation and Operation Guide Safety Warnings 6 ...
Page 10: ...ACCQPrep HP125 Table of Contents 4 ...
Page 16: ...ACCQPrep HP125 Section 1 Introduction 1 6 ...
Page 100: ...ACCQPrep HP125 Section 5 PeakTrak 5 36 ...
Page 128: ...ACCQPrep HP125 Section 6 Maintenance 6 28 ...
Page 130: ...ACCQPrep HP125 Appendix A A 2 This page is intentionally left blank ...
Page 132: ...ACCQPrep HP125 Appendix A A 4 This page is intentionally left blank ...
Page 134: ...ACCQPrep HP125 Appendix A A 6 This page is intentionally left blank ...
Page 135: ...A 7 Appendix A ACCQPrep HP125 This page is intentionally left blank ...
Page 136: ...A 8 This page is intentionally left blank ...
Page 138: ......
Page 140: ......
Page 142: ......