5.2.3 MS
The MS menu item is only available when a PurIon Mass
Spectrometer is added to the ACCQ
Prep
system. The following
options are available:
Method Development –
Choose this command to open the
MS M
ETHOD
D
EVELOPMENT
window to verify ionization
conditions for the compounds to be purified. This command
is not available from a remotely connected computer via a
web browser.
Ion Finder –
Enables you to find fragments and adducts
based on the molecular weight of their desired compound.
·
D
ESIRED
M
ASS
- enter the molecular weight of your
compound, rounded to the nearest integer.
·
D
ETECTED
I
ONS
- list of ions to be detected.
·
D
ETECTION
I
ONS
- list of possible adducts or
fragments based on the value entered in the desired
mass control and detected by the system. Checking
one of these values adds it to the D
ETECTED
I
ONS
list.
Manual Control –
Use this command to open the MS
M
ANUAL
C
ONTROL
window to prime and control the carrier
solvent pump and switching valve on the mass
spectrometer fluid interface. This command is not available
from a remotely connected computer via a web browser.
Ionization settings –
This command opens the I
ONIZATION
S
ETTINGS
window to allow creation of custom ionization
parameters that can be saved for various compound classes
and projects. This command is not available from a
remotely connected computer via a web browser.
Operate –
This command sets the status of the PurIon mass
spectrometer from standby to an operational mode by
applying power to the heaters and dynode detector. The
nitrogen gas is also supplied to the nebulizer. If the mass
spectrometer is already in operate mode, the S
TANDBY
button will display.
Standby –
This command is available when the PurIon is in
the O
PERATE
mode. It turns off all high voltages within the
PurIon including the dynode detector which prolongs its
life. In addition, it sets the inlet capillary heater
temperature to 50 °C and the remaining heaters off and
lowers the gas flow rate to ~.2 liters/minute. When in
standby mode, the operate command is used to return the
PurIon to the operational state (high voltages on, heaters
on, and gas flow at the proper rate of ~4 liters/min).
Shutdown –
This command removes power to the heaters,
and removes power from the turbomolecular pump. A
window will appear confirming that you want to initiate the
shutdown procedure. This command is used to shut down
the PurIon mass spectrometer for maintenance or to move
the system.
Summary of Contents for ACCQPrep HP125
Page 2: ......
Page 6: ...ACCQPrep HP125 Installation and Operation Guide Safety Warnings 6 ...
Page 10: ...ACCQPrep HP125 Table of Contents 4 ...
Page 16: ...ACCQPrep HP125 Section 1 Introduction 1 6 ...
Page 100: ...ACCQPrep HP125 Section 5 PeakTrak 5 36 ...
Page 128: ...ACCQPrep HP125 Section 6 Maintenance 6 28 ...
Page 130: ...ACCQPrep HP125 Appendix A A 2 This page is intentionally left blank ...
Page 132: ...ACCQPrep HP125 Appendix A A 4 This page is intentionally left blank ...
Page 134: ...ACCQPrep HP125 Appendix A A 6 This page is intentionally left blank ...
Page 135: ...A 7 Appendix A ACCQPrep HP125 This page is intentionally left blank ...
Page 136: ...A 8 This page is intentionally left blank ...
Page 138: ......
Page 140: ......
Page 142: ......