Section 2 Installation
2-18
2. Turn O
N
the power to the ACCQ
Prep
. A red light on the
AutoSampler will light up, the arm will move to the H
OME
position, and then the light will turn blue (Figure 2-29).
Figure 2-29 Power switch for the ACCQPrep
Prime AutoSampler Wash
Station
1. Ensure the wash station supply line is placed into a suit-
able solvent (typically the strong chromatographic solvent)
and the waste line is routed to a suitable container.
2. Go to T
OOLS
|A
UTOMATION
M
ANUAL
C
ONTROL
, and select
the S
TART
W
ASH
button (Figure 2-30).
Figure 2-30 Screen of Start Wash button
3. Ensure the wash solvent is pumped into the center portion
of the wash station and when full, the fluid drains into the
overflow drain (Figure 2-31).