Section 2 —475
OPERATING INFORMATION
Introduction
This section of the manual is intended to allow the
operator to become familiar with the instrument's power
requirements, functions of controls and connectors, and
how to obtain a few basic displays. For more complete
operating
inform ation, refer to the 475 Operators
Handbook.
Operating Voltage
This instrument is designed fo r operation from a
power source with its neutral a t or near earth
(ground) potential with a separate safety-earth con
ductor. I t is n o t intended fo r operation from two
phases o f a multi-phase system
,
or across the legs o f a
single-phase three-wire system.
This instrument can be operated from either a 115-volt
or 230-volt nominal line voltage source, 48 to 440 hertz.
The Line Voltage Selector switch in the instrument
converts the instrument from one nominal operating
voltage to the other. The Regulating Range Selector
assembly on the instrument rear panel selects one of three
regulating ranges for each nominal line voltage, and also
contains the line fuses for overload protection for both
nominal line voltages. To select the correct nominal line
voltage and regulating range, proceed as follows:
1. Disconnect the instrument from the power source.
2.
To convert from 115-volts nominal to 230-volts
nominal line voltage, set the selector switch to the 230 volts
position (toward the rear of the instrument). Change the
line-cord plug to match the power source or use a
115-to-230 volt adapter.
NOTE
Color-coding o f the cord conductors is as follows (in
accordance with National Electrical Code):
Line
Black
Neutral
White
Safety earth (ground)
Green (or green with
yellow tracer)
3.
To change regulating ranges, loosen the two captive
screws which hold the cover onto the regulating range
selector assembly; then pull to remove the cover.
4.
Pull out the range selector switch bar (see Fig. 2-1).
Slide the bar to the desired position and plug it back in.
Select a range which is centered about the average line
voltage to which the instrument is to be connected (see
Table 2-1).
TABLE 2-1
Regulating Ranges
Regulating Range
Range Selector
Switch Position
115-Volts
Nominal
230-Volts
Nominal
LO (switch bar in
upper holes)
99 to 121 volts
198 to 242 volts
M (switch bar in
middle holes)
104 to 126 volts
208 to 252 volts
HI (switch bar in
lower holes)
108 to 132 volts
216 to 264 volts
Fig. 2-1. Power supply regulating range selector.
2-1
Summary of Contents for 475
Page 6: ...475 475 Oscilloscope ...
Page 20: ......
Page 46: ......
Page 151: ...A B C D E F G Fig 7 4 P 0 A4 Vertical Mode Switch circuit board ...
Page 152: ...n m m m 1 t siumm c ii im ...
Page 154: ...Diagrams 475 A B C D E F G ...
Page 155: ...GRID LOCATOR Fig 7 8 P O A3 Vertical Preamp circuit board ...
Page 157: ...Diagrams 475 ...
Page 158: ...GRID LOCATOR 2 3 4 5 6 7 8 9 1 10 11 1 12 Fig 7 10 P 0 A3 Vertical Preamp circuit board ...
Page 160: ...Diagrams 475 GRID LOCATOR Fig 7 11 A5 Vertical Output circuit board ...
Page 163: ...Diagrams 475 GRID LOCATOR Fig 7 12 P O A8 Trigger Generator And Sweep Logic circuit board ...
Page 167: ... 415 OSCILLOSCOPE B T R IG G E R G E N E R A T O R g B TR IG G E R GEN ER A TO R ...
Page 168: ...Diagrams 475 GRID LOCATOR ...
Page 172: ...SWEEP G E N ER A TO R S ...
Page 175: ...TIM IN G HORIZONTAL DISPLAY SWITCHING ...
Page 177: ...A 011 HORIZONTAL AMPLIFIER 8 l j k J Ll J Ol Z o N C O X U 1 C L O o v O o _J V v P O in r ...
Page 179: ... 476 OSCILLOSCOPE ...
Page 180: ...POWER SUPPLY ...
Page 182: ...CRT CIRCUIT ...
Page 185: ...C A L IB R A T O R A N D F A N C I R C U I T 4 C A LIB R A TO R FAN MOTOR C IR C U IT ...
Page 189: ......
Page 190: ...4 T 475 OSCILLOSCOPE FIG 1 FRO NT PANEL ...
Page 191: ...475 OSCILLOSCOPE FIG 2 MAIN FRAME ...
Page 199: ......
Page 200: ...FIG 3 RIGHT SIDE l 475 OSCILLOSCOPE ...
Page 201: ...V S X O t l B O T T OVJ 8 8 ...
Page 209: ......
Page 210: ...5 ...
Page 211: ...4 4 75 O S C ILLO S C O P E FIG 5 CAB IN ET ...
Page 215: ......
Page 237: ......
Page 241: ... 1 I R I II9A po S IT I o n f i n e I R l 130 PAGE 4 OF 5 R 1234 Ml 9 91 9 473 ...