Output Amp Settings
4-18
Output Amp Settings
In the
System
dialog, click the tab of the slot in which the output amp is installed to set the output ranges.
<AO-8> is shown at the
Slot
tab.
Output Range Setting
You can select the output range from 1 V to 5 V at 0.1 V steps.
z
In the recording at the low-speed sampling mode (1 kHz to 1/60 Hz), you cannot use the analog output
at the output amp.
z
You can playback the data recorded at the low-speed sampling mode (1 kHz to 1/60 Hz) by changing
the playback sampling frequency at 1.5 kHz or higher.
z
Make sure to specify the source slot installed the input amp at the
Output Slot
.
z
At the LX series, the AD value of +/- 25000 is correspondent to +/- 100 % of the input range. So that
the output amp outputs the signal at the scale of +/- 100% as the specified output range. For
example, if you 1V signal at the 1 V input range and you set the output range at 2 V, you will
2 V signal correspondent to the recorded 1 V signal.
Select the output ranges.
Specify a source slot of the
input amp for output.