Recording Synchronization
5-25
Recording Synchronization
This section explains the optional recording synchronization function of the LX multiple units.
The LX Navi sets a master unit and up to 3 slave units connected with the recording synchronization cable
within the total of 10 m (as of November 2005) for recording synchronization recording.
z
The following PC (personal computer) system specifications are necessary for the recording
synchronization function.
OS:
Windows XP / Windows 2000
CPU: Pentium
Ⅳ
2.0GHz or better
Memory: 256Mbyte or more
Free HDD space: 2Gbyte or more
LX Navi version: V1.71H or later
。
z
If you use the IEEE1394 interface model of the LX and you need to update the LX Navi, update the device
driver to the latest (see Section 2).
z
You cannot use the IEEE1394 interface model and the Ethernet interface model mixed for the recording
synchronization.
z
When the recording synchronization function is enabled and LxNavi is started up, the reproducing
operation cannot be used. Before using the reproducing operation, re-launch LxNavi, disabling the
master/slave settings and
execute it as an independent unit.
z
Change the settings of the sampling frequency of recording and the recording media destination (Memory,
MO, PC card) to that of the master unit and slave unit. When controlling the master unit and slave unit
using a single PC, (simultaneously launching multiple units with LxNavi using recording synchronization)
set the directory name and file name of the file to be recorded, so they do not duplicate.
z
When controlling the master unit and slave unit using a single PC, limitations may occur in the operation
of the maximum sampling frequency and in the save function of Lx Navi settings content.
z
When using recording synchronization, launch LxNavi operations in order; master unit, slave 1, slave 2,
etc. When ending the program, end operations in the opposite order to which they were started and lastly
close the LxNavi master unit.
z
Use the optional LX
View
PL-S1001 wave monitor software or the AFC NEO PL-S1002 file converter
software to merge the multiple files recorded.
z
Keep minimum numbers to open the main windows of the slave units; otherwise the real time monitor
cannot be processed depending on the selected sampling frequency.