Input Amp Settings
4-5
Input Amp Settings
In the
System
dialog, click the tab of the slot in which the input amp is installed. The setting parameters are
valid depending on the type of the amps installed.
DC Amp AR-LXDC/DC2
<DC-8> is shown at the
Slot
tab.
The AR-LXDC2 is a DC input amp equipped the low-speed samplings (1 kHz to 1/60 Hz) addition to the functions
that the AR-LXDC supports.
See Section 6 for the specifications.
Range Settings
You can select from +/- 0.5, 1, 2, 5, 10, 20, or 50 V. A range of +/- 131% of the selected range can be
obtained for A/D (analog to digital) conversion value; however, the input margin level is approxi/-
120%. Input the signal, then click button on the right and specify the setting while checking the level.
Auto
When you input the signal and click
Auto
, the optimum input range is set automatically.
Same Setting
Check to apply the same settings to the channels later than the channel changed.
z
You cannot use the analog output signal of the monitor output and/or the analog output amp when
recording at the low-speed sampling (1 kHz to 1/60 Hz).
z
You can reproduce the data recorded at the low-speed sampling from the analog output amp by changing
the reproduce sampling speed at 1.5 kHz or higher.
Select the input range.
Check the input level.
Check the input levels of all channels.
Stop checking the input levels.
For the auto range setting.