6-6
Netra CT Server Service Manual • February 2007
3. You can only remove a board if
all
of the LEDs are in the following states:
■
Amber Okay to Remove LED on the system status panel is
on
■
Green Power LED on the system status panel is
off
■
Blue Hot-swap LED on the board is
on
Do not remove any board from the system if any of the LEDs are in a different
state from those listed here.
6.1.1.2
Rules and Restrictions for Hot-Swapping the Host Board
Following are the rules and restrictions you must understand before hot-swapping a
host CPU board:
1. You must have the alarm card and system controller board installed in the Netra
CT server before you can hot-swap the host CPU board.
2. When you remove a host CPU board from a server while the server is running,
the CompactPCI board is disabled, which means that the connection with all I/O
boards installed in the server is lost. Once the CompactPCI bus is disabled,
MCNet is disabled in the system.
After you install the host CPU board back into the server, CompactPCI bus is
enabled and attempts to connect to all occupied board slots.
Satellite CPU boards are unaffected by the disabling of the CompactPCI bus; they
continue to function as stand-alone processors when you hot-swap a host CPU
board.
3. If a host CPU is at the OpenBoot prompt (the
ok
prompt) and you want to hot-
swap the host CPU board, you must power off the host CPU slot through the
alarm card.
To power off the host CPU slot in a Netra CT server, access the alarm card and
through the alarm card command line interface enter:
Refer to the
Netra CT Server System Administration Manual
for more information
about accessing the alarm card.
Use the
poweron
command to power on the host CPU board after you install the
replacement host CPU board.
cli>
poweroff host
Summary of Contents for Netra CT 410
Page 16: ...xvi Netra CT Server Service Manual February 2007...
Page 20: ...xx Netra CT Server Service Manual February 2007...
Page 28: ......
Page 44: ...2 10 Netra CT Server Service Manual February 2007...
Page 48: ...3 4 Netra CT Server Service Manual February 2007...
Page 49: ...PART 2 Troubleshooting the System Troubleshooting the System Chapter 4...
Page 50: ......
Page 88: ......
Page 98: ...5 10 Netra CT Server Service Manual February 2007...
Page 138: ...6 40 Netra CT Server Service Manual February 2007...
Page 154: ...7 16 Netra CT Server Service Manual February 2007...
Page 176: ...8 22 Netra CT Server Service Manual February 2007...
Page 178: ......
Page 198: ...10 10 Netra CT Server Service Manual February 2007...
Page 199: ...PART 5 Illustrated Parts List Illustrated Parts List Chapter 11...
Page 200: ......
Page 208: ...11 8 Netra CT Server Service Manual February 2007...
Page 210: ......
Page 265: ...Appendix A Connector Pinouts A 55...
Page 266: ...A 56 Netra CT Server Service Manual February 2007...
Page 288: ...D 4 Netra CT Server Service Manual February 2007...
Page 296: ...Index 4 Netra CT Server Service Manual February 2007 W web sites third party xxvi...