Working with Alarm Outputs (J1)
Refer to the following pin-out information (
) and relay output details (
) when installing alarm output wiring.
NC = normally closed, NO = normally open
Nominal switching power: 1A 30 VDC, 0.5A 125 VAC (resistive load)
Break-Before-Make operation
I
SOLATION
L
EVELS
:
•Insulation resistance: Min. 1,000 MΩ @500 VDC
•Breakdown voltage:
•Between open contacts:750 Vrms for 1 min. (Detection current: 10 mA)
•Between contact and coil 1,000 Vrms for 1 min. (Detection current: 10 mA)
•Between contact sets 1,000 Vrms for 1 min. (Detection current: 10 mA)
Figure 3-2: Alarm Relay Output Detail
Table 3-1: Alarm Relay Outputs (J1)
Pin Number
Signal
Function
1
ALARM_1_NC
Relay 1 Normally Closed contact
2
ALARM_1_COM
Relay 1 Common contact
3
ALARM_1_NO
Relay 1 Normally Open contact
4
ALARM_2_NC
Relay 2 Normally Closed contact
5
ALARM_2_COM
Relay 2 Normally Common contact
6
ALARM_2_NO
Relay 2 Normally Open contact
7
ALARM_3_NC
Relay 3 Normally Closed contact
8
ALARM_3_COM
Relay 3 Normally Common contact
9
ALARM_3_NO
Relay 3 Normally Open contact
10
ALARM_4_NC
Relay 4 Normally Closed contact
11
ALARM_4_COM
Relay 4 Normally Common contact
12
ALARM_4_NO
Relay 4 Normally Open contact