2
.........................................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
53
SUEZ © 2022
S
YSTEM
D
ESCRIPTION
S
YSTEM
S
PECIFICATIONS
Power Requirements
Fuses
Normal Operating Environment
Ambient Temperature
Maximum Relative Humidity
Maximum Altitude
Inputs
Outputs
Installation/Overvoltage Category
Safety Certifications
Pollution Degree
Display
Size
Weight
IP Rating
Optional Wi-Fi
Industrial Communication Protocols
total organic carbon
0.03 to 2,500 ppb as TOC
± 5% of measurement; ±0.1 ppb, whichever is greater
± 1% of measurement; ± 0.03 ppb, whichever is greater
5.5–8.0 in order to meet accuracy specifications
Online, Online Averaged, Online Timed, Grab
3 minutes for continuous online measurements
300 ppb O3 continuous
0.25 mL/min
Minimum 50 mL/min
1-95 °C (34-203 °F) (withstands short-term steam exposure)
Up to 100 psig
Insensitive to organic heteroatoms
12 months
3 significant digits
Linear Range
¹
Accuracy
Precision
Analysis Modes
Analysis Time²
Ozone Compatibility
Internal Sample Flow Rate (nominal)
External Flow Rate
Sample Temperature
Sample Pressure
Interferences
TOC Calibration Stability
Display Readout
instrument
100–240 VAC, 70 W, 50/60 Hz
No user-replaceable fuses
Intended for indoor use only
5-40 °C (41-104 °F)
Up to 95%, noncondensing
3,000 m (9,843 ft)
Two isolated binary inputs
Serial (RS-232), one USB, three 4-20 mA, four alarms, one Ethernet
II (protects against transients present in Category II power)
CE, UKCA, ETL listed. Conforms to UL Std. 61010-1
Certified to CSA 22.2 No. 61010-1
2 (normally only non-conductive pollution)
Backlit 10.1”,1280 x 800, touchscreen display
H: 43.5 cm (17.1 in) W: 47.5 cm (18.7 in) D: 28.4 cm (11.2 in)
13.9 kg (30.6 lb) SS Sampler; 13.6 kg (30.0 lb) standard iOS
IP 55
802.11ac/a/b/g/n Dual band 2.4/5 GHz
Modbus TCP/IP
1
TOC is not affected by conductivity up to 800 μS/cm.
²
Time to first measurement is 10 minutes.
pH Range