Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
18
SUEZ © 2022
W
ARNINGS
E
NGLISH
Warnings
This symbol on the instrument indicates that the user should refer to the
manual for operating instructions.
The iOS System and vial ports contain sharp needles designed to
pierce the septa of sample vials. Do
not
put fingers or inappropriate
materials into the iOS System or vial port.
Water in the iOS system may be hot. Before inserting a vial into the iOS
system after operating in Online mode, slide the door open and wait
30 seconds to allow sample to completely drain. Inserting a vial before
draining can result in hot water spray projecting upward out of the iOS
system.
This symbol indicates the protective earth terminal (ground) for the
Analyzer.
This symbol indicates the European Union Directive 2012/19/EU for
waste electrical and electronic equipment (WEEE) requires disposing
the Analyzer separately from standard waste.
To avoid potentially dangerous shock,
before
opening the Analyzer,
stop any analysis and power off the operating system. Click the
Power
Off
P
button in the Analyzer’s firmware, then turn off the power switch
and disconnect from the power supply. Any operation requiring access
to the inside of the Analyzer without turning off power, including
installation of maintenance items, could result in injury.
To protect against accidental exposure to ultra-violet radiation, do not
operate the UV lamp outside of its protective housing. The UV lamp
housing may be hot. Use extreme caution whenever touching the UV
lamp housing.
Risk Group 3 Warning: UV emitted from this product. Avoid eye and
skin exposure to unshielded product.