W
ARNING
AND
E
RROR
D
ESCRIPTIONS
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
301
SUEZ © 2022
5000
Error
Calibration
Verification
The instrument is unable to calibrate. Please contact Technical
Support or your local service provider.
Occurs during calibration protocols when a calibration calculation
cannot be performed due to attempting to perform an invalid math
operation.
5001
Error
Calibration
Verification
Calibration measurement is out of range
Occurs during TOC calibration protocols, when no measurement is
returned for one or more vials.
5800
Error
DI Pump
No current in the DI Pump was detected.
Occurs when the DI pump is on and the Analyzer detects a current
< 120 mA.
5802
Error
DI Pump
High current in the DI Pump was detected.
Occurs when the Analyzer detects an unexpectedly high current in
the DI pump:
•
The DI pump is
on
and the current > 500 mA
•
The DI pump is
off
and the current > 120 mA
5804
Error
IC/TC Valve
No current was detected in the IC/TC Valve.
Occurs when the IC/TC valve is actuated and the Analyzer detects
a current < 5 mA.
5805
Error
IC/TC Valve
The current detected in the IC/TC Valve is unexpectedly high.
Occurs when the Analyzer detects:
•
> 200 mA when valve is actuated
•
> 5 mA when valve is not actuated
5807
Error
DI Loop Fan
High current in the DI Loop Fan was detected.
Occurs when the fan is on and the Analyzer detects a current
> 200 mA.
6800
Error
UV lamp
UV lamp failed.
6901
Error
IC sample pump
No current in the Sample Pump motor was detected.
Occurs when the pump is on and the Analyzer detects a current
< 60 mA
6902
Error
IC sample pump
High current in the Sample Pump motor was detected.
Occurs when the Analyzer detects:
•
> 1000 mA when the pump is on
•
> 60 mA when the pump is not on
6903
Error
Sample Pump
No motion on the Sample Pump motor was detected.
Occurs when the pump is running but no motion is detected with
quadrature encoder.
#
Warning/
Error
Name
Description