C
ONFIGURING
A
LARM
O
UTPUTS
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
152
SUEZ © 2022
•
Backflush (wizard) — Follow the screen instructions to
backflush of the sample side
of the DI side fluidics. This can clear any clogs in tubing when a lack of flow is indicated
by
Analyzer.
•
Binary Inputs Setup — Use the
Binary Inputs Setu
p screen to set alarm parameters.
Refer to
“Configuring Binary Input (Start/Stop)” on page 155
instructions.
•
Drain Instrument (wizard) — Follow the screen instructions to drain the sample side
and/or DI loop.
•
Ethernet Setup — Use the
Ethernet Setup
screen to enable an Ethernet
“Configuring an Ethernet Connection” on page 157
for step-by-
step instructions.
•
Flow Sensor Setup — Use the
Flow Sensor Setup
screen to enable the flow sensor
and the
No Flow
“Configuring the Flow Sensor” on page 158
step-by-step instructions.
•
Instrument Startup Wizard — (Follow these wizard screens
only
when first
installing the Analyzer, as described in
“Step 7: Install the DI Water System” on
) For instructions on filling the DI Water
Reservoir, priming the DI Loop, and rinsing down the measurement module
after
the initial installation process
, refer to
•
Level Sensor Setup — Use the
Level Setup
screen to enable the DI Reservoir
“Configuring the Flow Sensor” on page 158
instructions.
•
UV Lamp Setup — Use the
UV Lamp Setup
screen to temporarily turn off the UV
lamp. See
“To enable the UV Lamp” on page 161
for step-by-step instructions.
•
Printer Setup — Use the
Printer Setup
screen to enable a PDF printer connected to
the Analyzer directly or over an Ethernet network. Refer to
for step-by-step instructions.
•
WiFi Setup — Use the
WiFi Setup
screen to enable a WiFi connection. Refer to
“Configuring a WiFi Connection (Optional)” on page 164
for step-by-step
instructions.
Configuring Alarm Outputs
The Sievers M500e TOC Analyzer has four customizable alarm outputs,
configurable to trigger if the measured TOC, TC, or IC exceeds a set value — or if
a warning or error occurs. Use this instruction for configuring the alarms wired
earlier in
“Step 4: Install I/O Control Wiring (Optional)” on page 63
To set up alarm output
NOTE:
You may need to consult with your remote operations
center to determine some of the values.