![Suez Sievers M500e Operation And Maintenance Manual Download Page 148](http://html1.mh-extra.com/html/suez/sievers-m500e/sievers-m500e_operation-and-maintenance-manual_1389721148.webp)
U
PGRADING
F
IRMWARE
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
148
SUEZ © 2022
4.
Click inside the
Enter Activation Code
field and use the alphanumeric keypad
to enter the activation code provided by SUEZ. The feature appears in the
Features To Be Activated
area.
5.
To install the feature shown, click
Apply
. A confirmation message appears.
6.
Click
Yes
to continue. The Activations dialog box closes and the system adds
the feature shown to the Analyzer.
Upgrading Firmware
IMPORTANT:
System constants, Analyzer settings, and a Database Backup must be
saved before upgrading Analyzer firmware. See
“Saving Analyzer Settings” on page 144
, and
Use these instructions to install firmware updates:
•
“To upgrade firmware using a USB memory device” on page 148
•
“To upgrade firmware through a network connection” on page 150
To upgrade firmware using a USB memory device
1. Power cycle the Analyzer and before re-starting, insert the USB memory
device (with the firmware update) into the USB port.
2.
Navigate to the
Analyzer Settings
panel (
Settings
s
screen >
Instrument
Settings
tab).