P
ROGRAMMING
THE
TOC A
UTOZERO
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
187
SUEZ © 2022
P
ROGRAMMING
THE
TOC A
UTOZERO
The TOC Autozero protocol determines the TOC offset, which is the difference
between the IC and TC measurements with the UV lamp off. It also determines the
background conductivity curves for each channel.
The Analyzer can be set to automatically perform a TOC Autozero at a specific
interval. The TOC Autozero is recommended only if the TOC concentration in the
sample stream is 50 ppb or less.
To program the TOC Autozero
1. If the Analyzer is taking measurements, click
Stop Analysis
.
2. On the
System Protocols
p
screen, select the
Calibration
tab.
3. Select
TOC Autozero
, then click
Setup
.
4. The
Setup TOC Autozero
dialog box appears.