Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
124
SUEZ © 2022
NOTE:
Changing the value for the time range on the X axis here
also changes the value for the time range on the Result
Preferences panel (Settings
s
screen > System Setup tab). In a
like manner, changing the value for the time range on the Result
Preferences panel also changes the value here. See
graph display options” on page 137
5.
Click
OK
. The trend graph reappears with the selected time range,
background, and point size. The example shows a trend graph with a time
range of 1 hour, a blue background, and 9-pixel data points.
Setting the Measurement Mode
The Sievers M500e TOC Analyzer provides three different
Online
modes for running the
Analyzer with a continuous water source. A
Grab
mode is also available for sampling
from a vial in the Analyzer’s Standard iOS System.
Select one of these modes as the default measurement mode using the
Protocols
p
screen. The Analyzer uses this measurement mode setting for analysis started from the
Home
h
screen (which also displays the active measurement mode in the
Protocols
field).
To revise the active measurement mode
1. On the
Protocols
p
screen, select the
User Defined Method
tab.