TC/IC C
ONDUCTIVITY
A
UTOZERO
P
ROTOCOL
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
219
SUEZ © 2022
5.7 Referencing the
Summary
tab, record the results data on the
TC/IC
Conductivity Autozero Worksheet
.
5.8 Do one of the following:
5.8.1 If the calibration
passed
, click
Apply
to accept the calibration. A
confirmation message appears. Click
Apply calibration
, then
Close
.
5.8.2 Click
Reject
to reject the new offset. A confirmation message appears.
Click
Reject calibration
, then
Close
.
5.9 To print or export the results, click the corresponding button.