Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
137
SUEZ © 2022
•
TC
4.
In the
Automatically Print Results
area, select one or more to automatically
print results:
•
Grab Protocols
•
System Protocols
•
Online Modes (If selected, also select one of the following options from the
Automatically Print Results
list:
every sample
,
every hour
,
every 2 hours
,
every 4 hours
,
every 8 hours
.)
Setting Trend Graph Display Options
Set the time scale and which data appears on the trend graph, which is available for
viewing on the
Home
h
screen and
Data
D
screen. Rejected Repetitions can also be
set whether to appear in printed and exported results.
To set trend graph display options
1. Navigate to the
Result Preferences
panel (
Settings
s
screen >
System Setup
tab).
2.
In the
Trend Graph Display Options
area, configure the following:
•
X Axis Time Range
list — Select one of the following options from the list:
1 hour,
4 hours
,
8 hours
,
12 hours
,
1 day
,
3 days
, or
one week
.
•
Y Axis Scaling
— Select
Auto
(using default minimum and maximum values)
or
Manual
(to set the minimum and maximum values).
•
Y1 Axis
— Select or de-select whether to include
Carbon (TOC, IC, TC)
. If the
manual Y-Axis scaling
option was previously selected, click inside each of
the
Min
and
Max
fields and use the alphanumeric keypad to set the
minimum and maximum value (TOC values in ppb).