![Suez Sievers M500e Operation And Maintenance Manual Download Page 174](http://html1.mh-extra.com/html/suez/sievers-m500e/sievers-m500e_operation-and-maintenance-manual_1389721174.webp)
A
RCHIVING
THE
D
ATABASE
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
174
SUEZ © 2022
7.
Specify the number of months of history data to retain in the Analyzer’s
database.
For example, on 15 December 2020, a user archives the database and retains
2
months of history data. The Analyzer retains all history data from
15 October 2020 through 15 December 2020. History data older than
15 October 2020 is removed from the Analyzer’s database. All history data
(including history data older than 15 October 2020) is included in the backup
file.
8.
Click
OK
. A progress bar appears. When the archive is complete, the backup
file appears in the
Available Backups to Restore
section. A message appears
stating that the database was successfully archived and the Analyzer’s’
firmware must be restarted.
9.
Click
OK
to restart the Analyzer’s firmware. When the initialization process is
complete, the
Home
h
screen appears.