I
NSTALLATION
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
106
SUEZ © 2022
7.
Leave the default file name or click inside the
File Name
field. Use the
alphanumeric keypad to type a new file name (but not a file type extension)
and click
OK
. The exported file(s) are assigned the file type extension(s)
selected in step
.
8.
In the
Name
area, select the folder to receive the exported file and click
Save
.
A confirmation message appears.
9.
Click
OK
to confirm and close the message.
10.
If a USB memory device was used, remove it for safe keeping.
11.
Save the exported Settings file. Technical Support (or your local service
provider) might need to reference the system settings for Calibration constants
and other key parameters.
Configuring the Data I/O
If alarm outputs, 4-20 mA outputs, and/or binary inputs were installed, configure the
connections with the appropriate values for those features now.
NOTE:
You may need to consult with your remote operations
center to determine some of the values.
To set up alarm output
The Analyzer has four customizable alarm outputs, configurable to trigger if the
measured TOC, TC, or IC exceeds a set value — or if a warning or error occurs.
Use this instruction for configuring the alarms wired earlier in
Control Wiring (Optional)” on page 63
1. On the
Settings
s
screen, select the
Hardware
tab.