R
INSING
THE
A
NALYZER
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
314
SUEZ © 2022
3.
(Optional) Select the
Start Online Analysis When Finished
option.
4.
Click
S
TART
. The Analyzer begins the rinse process.
5.
Upon completion of the rinse (approximately one hour), a “Accelerated Rinse
is complete.” message appears.
6.
Click
C
LOSE
.
7.
Open the Analyzer door and inspect for any leaks (especially around the DI
water reservoir).
8.
If a leak is detected, verify all fittings are tight and secure. Also, confirm the
water level in the DI water reservoir is full. If necessary, turn the Analyzer
power switch to
Off
and add more water using the water bottle (with filling
nozzle) from the Analyzer’s
Accessories
kit.
9.
Close and latch the door to the Analyzer case.
10.
Turn Analyzer power switch to the
On
position. The Analyzer is now ready to
take valid TOC measurements.