U
PGRADING
F
IRMWARE
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
151
SUEZ © 2022
4.
Enter the network path, user name, and password to access the network folder
and click
OK
. A screen that lists files within the network folder appears.
5.
Select the upgrade file and click the
Open
button. The Analyzer starts the
upgrade. After the progress indicator reaches 100% complete, a
Restart
button appears.
6.
Click
Restart
to complete the upgrade process. The
Home
h
screen
appears.
S
ETTINGS
S
CREEN
— H
ARDWARE
T
AB
The
Hardware
tab includes a scrollable menu of various Analyzer functionality for
Analyzer setup and other hardware features. Each menu selection either includes
wizard screens to guide the user with step-by-step instructions and photos or a settings
screen. For each settings screen, accompanying instructions are included in this
section.
For each menu item, select the Hardware feature in the scrolling table and click
Run
.
The corresponding wizard or settings screen appears.
The menu includes the following:
•
Accelerated Rinse (wizard) — Follow the screen instructions to perform an
accelerated rinse of the Analyzer.
•
Alarms Setup — Use the
Alarms Setu
p screen to set alarm parameters. Refer to
“Configuring Alarm Outputs” on page 152
for step-by-step instructions.
•
Analog Output Calibration (wizard) — Follow the screen instructions to calibrate
each analog output.
•
Analog Outputs Setup — Use the
Analog Outputs Setu
p screen to set 4-20mA
analog output parameters. Refer to