M
AINTENANCE
Replacing Consumables and Maintenance Items
To ensure optimum performance of the Analyzer, routinely replace Analyzer
consumables and maintenance items. Refer to
Table 8-1, “Replacement Schedule for
for the recommended maintenance schedule. The in-line
filter, UV lamp, resin cartridge, and sample pump heads must be purchased from
SUEZ.
When replacing items that affect analysis, such as the UV lamp, resin cartridge, or
pump heads, perform a verification of the calibration. Only re-calibrate if verification
indicates that a new calibration is necessary.
For additional assistance, contact SUEZ Technical Support or your local service
provider. Training is also available from SUEZ.
This chapter covers replacement instructions for all routine maintenance items that are
replaced on a regular schedule; instructions for non-routine maintenance items are
provided in those items’ packaging.
A one-year maintenance record log is provided for convenience to ensure that all
required maintenance is performed when necessary (see
Year Maintenance Worksheet,” on page 288
).