Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
154
SUEZ © 2022
Configuring 4-20 mA Analog Outputs
To set up 4-20 mA analog outputs
If alarm outputs, 4-20 mA outputs, and/or binary inputs were installed, configure
the connections with the appropriate values for those features now. The Sievers
M500e TOC Analyzer has three 4-20 mA outputs. Select the output range for the
4-20 mA analog outputs by following these steps. Instructions for wiring the 4-
20mA output can be found in Chapter 3:
1. On the
Settings
s
screen, select the
Hardware
tab.
2.
On the scrolling menu, select
Analog Outputs Setup
and click
Run
. The
Analog Outputs Setup
screen appears.