TOC M
ULTI
-
POINT
C
ALIBRATION
P
ROTOCOL
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
213
SUEZ © 2022
5.0 P
ROCEDURE
5.1 If the Analyzer is taking measurements, click
Stop Analysis
.
5.2 Export the current system settings for future reference or reloading by
following the steps in
"To save the system settings" on page 192
.
5.3 Back up the database and settings by following the steps in
database and settings" on page 193
.
5.4 On the
System Protocols
p
screen, select the
Calibration
tab.
5.5 Select
TOC Multi-point Calibration
, then click
Start
.
5.5.1 A
Please insert...
dialog box appears.
5.5.2 Open the door to the iOS System and wait 30 seconds for water to
drain.
5.5.3 Insert the
Cal Blank 1
vial into the iOS System and click
Vial ready->
Continue
. The Analyzer begins analysis of the vial.
5.5.4 When analysis of the
Cal Blank 1
vial is complete, a
Please insert Rw
Blank...
dialog box appears.