Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
284
SUEZ © 2022
12.
Install the resin cartridge into the
RUN
position. Simultaneously insert one
elbow-fitting end of the resin cartridge into the
RUN
port and the other elbow-
fitting end into the
DI Pump Outlet
port.
13.
Verify that the ends are secure before moving to the next step.
14.
Restore power to the Analyzer.
15.
Enter the installation date for the resin cartridge in the firmware, as described
in
“Entering New Consumables Installation Dates” on page 287
16.
Let the Analyzer run for an additional 30 minutes to circulate water in the
DI loop.
17.
Turn off the Analyzer and open the door to check and troubleshoot any leaks.
18.
Restore power to the Analyzer.
19.
Refer to
“Accuracy & Precision Verification Protocol” on page 223
. Only
re
-
calibrate if verification indicates that a new calibration is necessary.